Nuclear Science and Technology - Volume 9, Number 4, December 2019

Abstract: Grid spacers installed in subchannels of fuel assemblies for nuclear reactors can promote

heat transfer. However, the fluid velocity and bubble behavior are greatly affected as the crosssectional area of the flow passage changes. Therefore, the void fraction distribution behind the

obstacle that simulates the grid spacer shape simply was measured by using a wire mesh sensor

(WMS) system. Moreover, a two-phase flow analysis was performed to investigate the effect of the

obstacle on the bubble behavior in a vertical duct.

Keywords: Bubbler, Two-phase flow, Obstacle, Vertical duct, WMS, Experiment, Analysis.

I. INTRODUCTION

Clarifying two-phase characteristics

in a nuclear reactor core is important in

particular to enhance the thermo-fluid safety

of nuclear reactors. Moreover, correct data

on bubbly flow in subchannels with spacers

are needed in order to verify two-phase flow

models in conventional nuclear safety

analysis codes and validate predicted data

by current CFD codes like a direct twophase flow analysis code (Douce, et al.,

2010). Spacers installed in subchannels of

fuel assemblies have the role of keeping the

interval between adjacent fuel rods constant.

Similarly, in case of PWR the spacer has

also the role as the turbulence promoter.

When the transient event occurs in a nuclear

reactor, two-phase flow is generated by

boiling of water due to heating of fuel rods.

Therefore, it is important to confirm the

bubbly flow behavior around the spacer.

The purpose of this study is to make the

effect of the spacer affecting the bubbly

flow clear and obtain code validation data.

So bubble dynamics around the simply

simulated spacer was visually observed and

the void fraction and interfacial velocity

distributions just behind the simulated

spacer was measured.

II. EXPERIMENTAL METHOD

1. Experimental Apparatus

The experimental apparatus mainly

consists of a measuring section and

water/air supply lines and is shown in Fig.

1. The measuring section includes a vertical

flow channel with a diameter of 58 mm

made of an acrylic resin, inlet and outlet

plenums, and an air injection nozzle with

120 injection holes with a diameter of 0.6

mm. Water flows through the inlet plenum

into the flow channel and goes up through

the measuring section to the outlet plenum.

Air is supplied from the air injection nozzle

into the flow channel. As a result, water is

mixed with air and then a water-air twophase flow is formed as can be seen in Fig.

2. Bubbly flow conditions are determined

by both flow rates of water and air.

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 1

Trang 1

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 2

Trang 2

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 3

Trang 3

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 4

Trang 4

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 5

Trang 5

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 6

Trang 6

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 7

Trang 7

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 8

Trang 8

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 9

Trang 9

Nuclear Science and Technology - Volume 9, Number 4, December 2019 trang 10

Trang 10

Tải về để xem bản đầy đủ

pdf 60 trang xuanhieu 4160
Bạn đang xem 10 trang mẫu của tài liệu "Nuclear Science and Technology - Volume 9, Number 4, December 2019", để tải tài liệu gốc về máy hãy click vào nút Download ở trên

Tóm tắt nội dung tài liệu: Nuclear Science and Technology - Volume 9, Number 4, December 2019

Nuclear Science and Technology - Volume 9, Number 4, December 2019
e 
ionization chamber (Razor chamber). The 
output correction factor [ 
 ]
is 
a 
DO DUC CHI et al. 
51 
obtained from the tabulated output correction 
factors with respect to the machine specific 
reference field as below: 
[ 
 ]
[ 
 ]
[ 
 ]
 (4) 
III. RESULTS and DISCUSSION 
A. Output factors of collimation systems 
using TRS398 CoP 
Using the conventional formula from 
TRS398 CoP, we got the result as Table I. 
Table I. Output factors of collimation systems using TRS398 CoP and Razor chamber. 
Cone (mm) 4 5 7.5 10 12.5 15 17.5 100 (square) 
6X 0.420 0.523 0.662 0.746 0.797 0.834 0.859 1 
6XFFF 0.473 0.574 0.702 0.772 0.816 0.846 0.866 1 
MLC FS (mm) 0.5 1 2 3 4 5 7 10 
6X 0.529 0.754 0.861 0.895 0.921 0.940 0.968 1 
6XFFF 0.560 0.782 0.876 0.909 0.932 0.948 0.975 1 
Jaw FS (mm) 0.5 1 2 3 4 5 7 10 
6X 0.345 0.704 0.850 0.888 0.914 0.936 0.967 1 
6XFFF 0.377 0.736 0.867 0.906 0.929 0.947 0.974 1 
B. Output factor of collimation systems 
using TRS483 CoP: 
Intermediate field ( ) of 4 × 4 
cm
2
 was selected for calculation of jaw-
shaped fields and MLC-shaped fields. For 
cone, intermediate field was 17.5 mm 
conical field because we need to 
normalize these data to that of 10 × 10 
cm
2
 field size. The results were obtained 
as Table II. 
Table II. Output factor of collimation systems using TRS483 CoP (Razor chamber and Razor diode) 
Cone (mm) 4 5 7.5 10 12.5 15 17.5 100 (square) 
Square Equi. Field 
size (cm) 0.708 0.885 1.327 1.77 2.212 2.655 3.097 10 
6X 0.522 0.599 0.713 0.773 0.814 0.841 0.864 1 
6XFFF 0.578 0.648 0.745 0.797 0.830 0.856 0.871 1 
MLC Field size(mm) 0.5 1 2 3 4 5 7 10 
6X 0.608 0.774 0.871 0.905 0.927 0.945 0.971 1 
6XFFF 0.643 0.792 0.881 0.918 0.938 0.954 0.978 1 
Jaw Field size(mm) 0.5 1 2 3 4 5 7 10 
6X 0.619 0.756 0.84 0.876 0.901 0.924 0.961 1 
6XFFF 0.652 0.771 0.846 0.884 0.907 0.928 0.963 1 
C. Comparison of results between TRS483 
and TRS398 CoP: 
Based on these results, the difference 
between ROF curves is significant between the 
two different methods (CoPs) for MLC-shaped 
field size and for jaw-shaped field size less than 
3 × 3 cm for both the 6X and 6XFFF beams. 
The smallest difference was observed 
with MLC-shaped fields while the biggest 
difference was observed with cone-shaped 
RELATIVE OUTPUT FACTORS OF DIFFERENT COLLIMATION SYSTEMS IN TRUEBEAM 
52 
fields as seen in Fig.3 and Fig.4. At 0.5 × 0.5 
cm
2
 squared field and 4 mm conical field, the 
output factor difference of 6X and 6XFFF 
beams were -44.2%/-42.2%, -13.0%/-13.0%, -
19.6%/-18.2% for jaw-shaped, MLC-shaped 
and cone-shaped fields, respectively. TRS398 
CoP gave underestimation of a relative output 
factor in comparison with TRS483 CoP. The 
large difference was always seen at field sizes 
smaller than 4 × 4 cm
2
. 
Fig. 3. Difference of TRS398 and TRS483 CoP in relative output factor of MLC and Jaws collimations. 
Fig. 4. Difference of TRS398 and TRS483 CoP in relative output factor of cone collimations. 
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F 
Jaw Field Size (cm x cm) 
6X Jaw Output Factors 
TRS483
TRS398
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F 
Jaw Field Size (cm x cm) 
6XFFF Jaw Output Factors 
TRS483
TRS398
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F 
MLC Field Size (cm x cm) 
6X MLC Output Factors 
Razor Chamber -
TRS398
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F 
MLC Field Size (cm x cm) 
6XFFF MLC Output Factors 
Razor Chamber -
TRS398
TRS483
0.4
0.5
0.6
0.7
0.8
0.9
1
0 10 20 30 40 50 60 70 80 90 100
R
O
F 
Coned Field Size (mm) 
6X Cone Output Factors 
TRS483
TRS398
0.4
0.5
0.6
0.7
0.8
0.9
1
0 10 20 30 40 50 60 70 80 90 100
R
O
F 
Coned Field Size (mm) 
6XFFF Cone Output Factors 
TRS483
TRS398
DO DUC CHI et al. 
53 
Noticingly, the Razor chamber’s 
reading differences between 10 × 10 cm
2
MLC-shaped field and 10 × 10 cm
2
 jaw-
shaped field were just 0.61% and 0.25% for 
6X and 6XFFF, respectively. Therefore, these 
relative output factors could be used for direct 
comparison between jaw-shaped field and 
MLC-shaped field of “the same” nominal 
field size. 
D. Comparison of results between 6X, 
6XFFF (TRS483 CoP): 
For the same collimation system, output 
factor comparisons were also made for 6X and 
6XFFF beams after applying TRS483 CoP 
(Fig. 5). The biggest differences in output 
factor were seen at 0.5 × 0.5 cm
2
 jaw-shaped 
field, 0.5 × 0.5 cm
2
 MLC-shaped field and 
4mm cone-shaped field with values of 5.3%, 
5.8% and 10.5%, respectively. 
Fig. 5. Difference in output factor of 6X and 6XFFF beams in each collimation system. 
E. Comparison of Output Factor curves 
between different collimation systems 
(TRS483 CoP): 
The relative output factor comparisons 
were made between MLC, jaws and Cone 
systems for both 6X and 6XFFF beams. 
Conical collimators are independent 
from MLC and Jaws systems. The conical 
collimation system has smallest relative 
output factor in comparison with that of MLC 
and Jaws systems for both 6X and 6XFFF 
beams as Fig. 6. 
For field sizes bigger than 1 × 1 cm
2
, jaw 
system has lower relative output factor than 
MLC’s but it is inverse for field size less than 1 
× 1 cm
2
. 
In a multi-centre analytical study of 
small field output factor calculations in 
radiotherapy reported by Krzysztof Chełmiński 
and Wojciech Bulski, for 2 × 2 cm
2
 MLC-
shaped fields of Varian linacs, the differences 
between the treatment planning system output 
factors (based on collected beam data) often 
exceeded 5% and were below 10% [11]. In 
our study, these differences were -1.1% (6X) 
and -0.5% (6XFFF) for MLC-shaped fields, 
1.3% (6X) and 2.6% (6XFFF) for jaw-shaped 
fields, -7.0% (6X) and -5.7% (6XFFF) for 
cone-shaped fields. The smaller differences 
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F 
Jaw Field Size (cm) 
TRS483 Jaws ROF 
6X
6XFFF
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F 
MLC Field Size (cm) 
TRS483 MLC ROF 
6X
6XFFF
0.5
0.6
0.7
0.8
0.9
1
2.5 5 7.5 10 12.5 15 17.5
R
O
F 
Coned Field Size (mm) 
TRS483 Cone ROF 
6X
6XFFF
RELATIVE OUTPUT FACTORS OF DIFFERENT COLLIMATION SYSTEMS IN TRUEBEAM 
54 
observed in our study for MLC-shaped field 
may came from our small field detector, the 
Razor chamber. 
A multinational audit of small field 
output factors calculated by treatment 
planning systems used in radiotherapy, the 
ROF for small fields calculated by TPSs 
were generally larger than measured 
reference data. On a national level, 30% and 
31% of the calculated ROF of the 2 × 2 cm
2
field exceeded the action limit of 3% for 
nominal beam energies of 6 MV and for 
nominal beam energies higher than 6 MV, 
respectively [12]. 
The discrepancy above may come 
from accuracy of treatment planning 
algorithms on measured output factors, 
especially for small fields. 
Fig. 6. Difference in output factor of difference collimation system for 6X and 6XFFF beams. 
CONCLUSIONS 
An international code of practice for the 
dosimetry of small static fields used in external 
beam radiotherapy (TRS483 CoP) was 
successfully applied to recalculate relative 
output factors for cone system with correction. 
Relative output factors for jaw collimation 
system were extensively obtained for field size 
less than 3 × 3 cm
2
 for Eclipse v.13.6 for 6X 
and 6XFFF beams using TRS483 CoP. 
Relative output factors were also measured for 
MLC collimation system to be compared with 
that of the jaw collimation system. The 
discrepancy of output factor between jaw-
shaped fields and MLC-shaped fields suggests 
that jaw-based beam data itself may not 
suitable for MLC-based treatment planning. 
Additional measurement of small beam 
percentage depth dose and profiles as well as 
specific modelling of photon beam for MLC 
system may be required. 
REFERENCES 
[1]. P. Andreo, “The physics of small megavoltage 
photon beam dosimetry,” Radiother. Oncol., 
vol. 126, no. 2, pp. 205–213, 2018. 
[2]. International Atomic Energy Agency, TRS483, 
“An International Code of Practice for 
Reference and Relative Dose Determination”, 
November, 2017. 
0.5
0.55
0.6
0.65
0.7
0.75
0.8
0.85
0.9
0.95
1
0 2 4 6 8 10
R
O
F 
Field Size (cm, square equivalent) 
6X TRS483 Output Factors 
MLC
Jaws
Cone
0.5
0.55
0.6
0.65
0.7
0.75
0.8
0.85
0.9
0.95
1
0 2 4 6 8 10
R
O
F 
Field Size (cm, square equivalent) 
6XFFF TRS483 Output Factors 
MLC
Jaws
Cone
DO DUC CHI et al. 
55 
[3]. I. J. Das, G. X. Ding, and A. Ahnesjö, “Small 
fields: Nonequilibrium radiation dosimetry,” 
Med. Phys., vol. 35, no. 1, pp. 206–215, 2008. 
[4]. I. J. Das, J. Morales, and P. Francescon, “Small 
field dosimetry: What have we learnt?,” in AIP 
Conference Proceedings, 2016. 
[5]. K. Park, J. Bak, S. Park, W. Choi, and S. W. 
Park, “Determination of small-field correction 
factors for cylindrical ionization chambers 
using a semiempirical method,” Phys. Med. 
Biol., vol. 61, no. 3, 2016. 
[6]. H. Bouchard, J. Seuntjens, S. Duane, Y. Kamio, 
and H. Palmans, “Detector dose response in 
megavoltage small photon beams. I. Theoretical 
concepts,” Med. Phys., vol. 42, no. 10, 2015. 
[7]. E. Mainegra-Hing and B. R. Muir, “On the 
impact of ICRU report 90 recommendations on 
kQ factors for high-energy photon beams,” Med. 
Phys., vol. 45, no. 8, pp. 3904–3908, 2018. 
[8]. H. Benmakhlouf, J. Sempau, and P. Andreo, 
“Output correction factors for nine small field 
detectors in 6 MV radiation therapy photon 
beams: A PENELOPE monte carlo study,” 
Med. Phys., vol. 41, no. 4, 2014. 
[9]. I. J. Das, M. B. Downes, A. Kassaee, and Z. 
Tochner, “Choice of Radiation Detector in 
Dosimetry of Stereotactic Radiosurgery-
Radiotherapy,” J. Radiosurgery, vol. 3, no. 4, 
pp. 177–186, 2000. 
[10]. M. Constantin, H. Hsu, S. Mansfield, and M. 
Svatos, “Mitigation Strategies to Reduce the 
Relative Extrafocal Dose ( RED ),” vol. 33, p. 
2525, 2006. 
[11]. K. Chełmiński and W. Bulski, “A multi-centre 
analytical study of small field output factor 
calculations in radiotherapy,” Phys. Imaging 
Radiat. Oncol., vol. 6, no. September 2017, pp. 
1–4, 2018. 
[12]. W. Lechner et al., “A multinational audit of 
small field output factors calculated by 
treatment planning systems used in 
radiotherapy,” Phys. Imaging Radiat. 
Oncol., vol. 5, no. September 2017, pp. 58–
63, 2018. 
[13]. Aspradakis, M. M. (2009). Small Field MV 
Photon Dosimetry. World Congress on 
Medical Physics and Biomedical Engineering, 
September 7 - 12, 2009, Munich, Germany, 
854–854. 
[14]. Absorbed Dose Determination in External 
Beam Radiotherapy: An International Code of 
Practice for Dosimetry based on Standards of 
Absorbed Dose to Water. Technical Reports 
Series No. 398, International Atomic Energy 
Agency, Vienna, 2000. 
INSTRUCTIONS FOR AUTHORS 
GENERAL INFORMATION 
Nuclear Science and Technology (NST), an 
international journal of the Vietnam Atomic 
Energy Society (VAES) and Vietnam Atomic 
Energy Institute (VINATOM), quarterly publishes 
articles related to theory and application of nuclear 
science and technology. All papers and technical 
notes will be refereed. 
It is understood that the paper has been neither 
published nor currently submitted for publication 
elsewhere. The copyright of all published papers 
and notes will be transferred in VAES. 
DETAILED FIELDS 
NST coves all fields of nuclear science and 
technology for peaceful utilization of nuclear 
energy and radiation. Authors should choose one 
of the following fields at the time they submit 
their manuscript: 1) Nuclear Physics, 2) Nuclear 
Data, 3) Reactor Physics, 4) Thermal Hydraulics, 
5) Nuclear Safety, 6) Nuclear I&C, 7) Nuclear 
Fuel and Materials, 8) Radioactive Waste 
Management, 9) Radiation Protection, 10) 
Radiation Technology, 11) Nuclear Techniques in 
Food and Agriculture, 12) Nuclear Medicine and 
Radiotherapy, 13) Nuclear Techniques in 
Industries, 14) Environment Radioactivity, 15) 
Isotope Hydrology, 16) Nuclear Analytical 
Methods, 17) Health Physics, 18) Fusion and 
Laser Technology. 
MANUSCRIPT SUBMISSION 
Manuscript for publication should be submitted to 
the Editorial Office in triplicate by postal mail. 
For electronical submission use 
nuscitech@vinatom.gov.vn. 
Submission Address 
Department of Planning, R&D Management 
Vietnam Atomic Energy Institute, 59 Ly Thuong 
Kiet Street, Hanoi, Vietnam 
E-mail: nuscitech@vinatom.gov.vn. 
MANUSCRIPT PREPARATION 
Manuscripts must be written in English with 
adequate margins and indented paragraph. All 
manuscript must use SI (metric) units in text, 
figures, and tables. Manuscripts should in general 
be organized in the following order: title, names of 
authors and their complete affiliation including zip 
code, abstract (not exceeding 200 words), 
keywords (up to 7), introduction, main body of a 
paper, acknowledgments, references, appendices, 
table & figure captions, tables and figures. 
Unnecessary sections may be omitted. 
Headings: Use I, II, for major headings and A, 
B,  for secondary headings. 
Mathematical formulas: All mathematical 
formulas should be clearly written, with special 
consideration to distinctive legibility of sub-and 
superscripts. Equation (at least the principal ones) 
should be numbered consecutively using Arabic 
numerals in parentheses in the right hand margin. 
Tables and Figures: Tables should be numbered 
with Roman numerals. Figures should be 
numbered consecutively with Arabic numerals in 
order of their first appearance and have a complete 
descriptive title. They should be typed on separate 
sheets. Tables should no repeat data which are 
available elsewhere in the paper. Figures should 
be original ink drawing or computer drawn figures 
in the original and of high quality, ready for direct 
reproduction. Figures should be referred to in the 
text as, for example, Fig. 1., or Fig. 2. . 
Reference: References should be listed at the end 
of the text and presented as follows: 
[1] C. Y. Fu et al., Nuclear Data for Science and 
Technology, S. M. Qaim (Ed.), p. 587 (1991). 
[2] C. Kalbach, Z. Phys, A283, 401 (1977). 
[3] S. Shibata, M. Imamura, T. Miyachi and M. 
Mutou, “Photonuclear spallation reactions in 
Cu”, Phys. Rev. C 35, 254 (1987). 
KHOA HỌC VÀ CÔNG NGHỆ HẠT NHÂN
Chịu trách nhiệm xuất bản 
TRẦN HỮU PHÁT 
Chịu trách nhiệm nội dung 
TRẦN HỮU PHÁT 
TRẦN CHÍ THÀNH 
Trình bày 
LÊ THÚY MAI 
In 200 cuốn, khổ 19x26,5cm tại Công ty TNHH Trần Công 
Địa chỉ: số 12 ngách 155/176 Đường Trường Chinh, Hà Nội 
Giấy đăng ký kế hoạch xuất bản số: 770/GP-BTTTT cấp ngày 20 tháng 5 năm 2011 
In xong và nộp lưu chiểu Quý I năm 2020 
25 000đ 

File đính kèm:

  • pdfnuclear_science_and_technology_volume_9_number_4_december_20.pdf